ZXK-200型纸张吸水率测定仪产品应用解释书

一、型号含义及临盆根据
A, model meaning and production basis
本仪器根据QB/T1662《克列姆接收性测定仪》所规定的各项技巧参数和技巧请求临盆的。
The instrument according to the requirements specified in the QB/T1662 absorption tester for "Klemm" of the technical parameters and production technique.
型号含义:
ZXK-200纸张吸水率测定仪
Type meaning:
Z XK 200
Z XK 200
最大年夜吸水高度(mm)
The maximum height (mm)
吸水率
Water absorption rate
纸张实验机
Paper testing machine
2、用处及应用范围
Two, the use and the use scope
本仪器实用于未施胶的纸和纸板毛细吸液高度的测定。(不适于10min内毛细高度小于5mm的纸和纸板)。
The utility model is suitable for the unsized paper and board Determination of capillary rise. (not suitable for 10min capillary height less than 5mm of paper and cardboard).
3、重要技巧参数
Three, the main technical parameters
1、丈量范围:5-200mm
1, measurement range: 5-200mm
2、试样尺寸:250x15mm
2, the sample size: 250x15mm
3、刻尺分度值:1mm
3, scale division value: 1mm
4、同时测定试样数量:10条
Simultaneous determination of 4, the number of samples: 10
5、刻尺数量:5
5, scale quantity: 5
四、仪器构造
Four, the structure of the instrument
仪器由底座、支柱、横梁、横板、刻尺、手柄及水槽构成(见图1)转着手柄起动偏爱机构使横梁带动横板及刻尺安稳上升或降低。
The instrument is composed of a base, a pillar, beam, plate, scale, handle and a water tank (see Figure 1) to rotate the handle starting eccentric mechanism to drive plate and beam scale steady rise or fall.
五、操作步调
Five, operation steps
1、迁移转变偏爱手柄使刻尺降到最低地位。
1, the rotation of the eccentric handle to scale down to the lowest position.
2、在水槽中注入过量的水,使一切刻尺的零点接触水面,经过过程调理支足来完成。
2, in the sink into the amount of water, so that all the scale of zero contact surface, is realized by adjusting the supporting foot.
3、转着手柄将刻尺升到最洼地位。
3, turn the handle will scale up to the highest position.
4、平放试样,在距试样一端205mm处整洁条与长度偏向垂直的直线,然后将10条试样夹在横
4, flat specimen, in a straight line from the sample is 205mm and the length of a vertical direction, and then the 10 specimens caught in the cross
板上刻尺的两侧,所划横线应与刻尺200mm刻线对齐。
On both sides of the plate scale, the underline and scale 200mm reticle alignment.
5、迁移转变偏爱手柄使刻尺降至最低地位,此时各刻尺零点恰好接触水面,而试样则深刻水中5mm这时候按动秒表计时,从而测定出试样毛细吸液的必定高度所需时间或一准时间内试样毛细吸液的高度值。普通采取10min±3s内的吸液高度,用s/mm或mm/s表示。
5, the rotation of the eccentric handle to scale to the lowest position, the scale zero just contact surface, and the samples into the water when the 5mm stopwatch, which measured a certain height sample capillary imbibition time required or within a certain period of time the sample capillary liquid absorbing height. Generally use the 10min ± 3S within the liquid height, denoted by s/mm or mm/s.
6、在读取数字时精确至1mm。如吸液润湿线倾斜或曲折,应按均匀高度读取成果。
6, in reading numbers to 1mm. Such as liquid wetting line inclined or curved, should read the results according to the average height.
7、深刻液体中的试样长度,可按产品标准或其它请求恰当延长,但要在实验申报中注明。
The length of specimen 7, deep in the liquid, can according to product standards or other requirements appropriately extended, but be noted in the test report.
六、仪器的考验
Six, test instrument
1、刻尺与横板应处于垂直地位。
1, scale and the transverse board should be in a vertical position.
2、五条刻尺的零点应位于同一程度线上。
2, five minute scale zero should be located on the same horizontal line.
3、偏爱轮升、降机构应灵活,操作轻松。
3, an eccentric wheel rise, delet mechanism should be flexible, easy operation.
4、试样夹子应保持稳定地夹紧试样
4, the specimen should be firmly clamped specimens
7、保养
Seven, maintenance
1、保持仪器干净,测试停止后应擦净加罩。
1, keep equipment clean, after the test should be cleaned with cover.
2、轴、套之间、定位键与槽之间定期注入光滑油。
2, shaft, sleeve, positioning key and the groove between the injected regularly into the lubricating oil.



装 箱 单

编号: 0 6 2
产品型号: Z X K— 200
产品称号: 纸张吸水率测定仪

序号 名 称 单位 数量 备注
1 主 机 台 1
2 水 盘 个 1
3 产品解释书 本 1
4 合 格 证 份 1